Hardware Failure as a Rare Complication of Percutaneous Coronary Intervention: A Case Report

Abstract

Percutaneous coronary intervention (PCI) is increasingly employed in the treatment of complex coronary artery disease. The entrapment or fracture of a coronary angioplasty guidewire is a rare complication of PCI. We herein describe a 61-year-old man who presented with chronic stable angina. The patient’s coronary angiogram revealed triple-vessel coronary artery disease, and he was scheduled for primary PCI. During the procedure, the guidewire fractured within the right coronary artery. Despite multiple attempts, the wire could not be retrieved. The wire unraveled in its coils, and its stretching resulted in its eventual snapping in the right radial artery. The initial plan was to attempt guidewire retrieval through a brachial cut-down, and if successful, to manage the obtuse marginal lesion by PCI, thereby precluding general anesthesia and a sternotomy. Unfortunately, the guidewire snapped at the brachial level, necessitating its retrieval by coronary artery bypass surgery. The patient remained asymptomatic and event-free over 6 months of follow-up.

1. Holmes DR, Jr, Williams DO. Catheter-based treatment of coronary artery disease: past, present, and future. Circ Cardiovasc Interv 2008;1:60-73.
2. Balbi M, Bezante GP, Brunelli C, Rollando D. Guide wire fracture during percutaneous transluminal coronary angioplasty: possible causes and management. Interact Cardiovasc Thorac Surg 2010;10:992-994.
3. Al-Moghairi AM, Al-Amri HS. Management of retained intervention guide-wire: a literature review. Curr Cardiol Rev 2013;9:260-266.
4. Balbi M, Bezante GP, Brunelli C, Rollando D. Guide wire fracture during percutaneous transluminal coronary angioplasty: possible causes and management. Interact Cardiovasc Thorac Surg 2010;10:992-994.
5. Singh J, Thingnam SK, Das D, Singh H, Sharma R, Vijayvergia R. Surgical removal of entrapped and broken percutaneous transluminal coronary angioplasty balloon catheter. Interact Cardiovasc Thorac Surg 2007;6:94-96.
6. Shimony A, Joseph L, Mottillo S, Eisenberg MJ. Coronary artery perforation during percutaneous coronary intervention: a systematic review and meta-analysis. Can J Cardiol 2011;27:843-850.
7. Ragkousis GE, Curzen N, Bressloff NW. Simulation of longitudinal stent deformation in a patient-specific coronary artery. Med Eng Phys 2014;36:467-476.
8. Vijayvergiya R, Swamy AJ, Grover A. Buddy wire technique: a simple technique for treating calcified lesion during percutaneous coronary intervention: a case report. J Invasive Cardiol 2006;18:E129-130.
9. Cunnington M, Egred M. GuideLiner, a child-in-a-mother catheter for successful retrieval of an entrapped rotablator burr. Catheter Cardiovasc Interv 2012;79:271-273.
10. Kaneda H, Saito S, Hosokawa G, Tanaka S, Hiroe Y. Trapped rotablator: Kokesi phenomenon. Catheter Cardiovasc Interv 2000;49:82-84.
11. Hartzler GO, Rutherford BD, McConahay DR. Retained percutaneous transluminal coronary angioplasty equipment components and their management. Am J Cardiol 1987;60:1260-1264.
Files
IssueVol 16 No 4 (2021): J Teh Univ Heart Ctr QRcode
SectionCase Report(s)
DOI https://doi.org/10.18502/jthc.v16i4.8603
Keywords
Coronary artery disease Percutaneous coronary intervention Catheterization Foreign bodies Coronary artery bypass

Rights and permissions
Creative Commons License This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
How to Cite
1.
PapaRao S, Bathala MR, Poddar A, Raman Ravindra kumar K, Sudhakar Paulraj Ignatius C, Padmanabhan C. Hardware Failure as a Rare Complication of Percutaneous Coronary Intervention: A Case Report. J Tehran Heart Cent. 2021;16(4):169-173.